Need help choosing?
Speak to one of our expertly trained specialists!
Need help Financing?
Check out financing options from OnePlace Capital.
All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.
This product has been discontinued by the manufacturer, though DRE may still be able to provide refurbished units. Please contact your Account Executive for more information on availability.
DRE offers the new Respironics Novametrix CO2SMO Veterinary CO2 Monitor and Pulse Oximeter.
Veterinary Capnograph
Veterinary Pulse Oximeter
Capnograph
Principle of Operation: Solid state mainstream sensor. Single beam, non-dispersive infrared absorption, ratiometric measurement.
Initialization Time: Capnogram within 15 sec. | full specifications within 60 sec.
Calibration: Routine calibration is not required. 15 sec. adapter zero performed when changing to different style of adapter.
Capnostat Mainstream CO2 Sensor:
Airway Adapters: Adult/pediatric reusable or single patient use, neonatal reusable, and sampling adapter for non-intubated patients.
Sampling Flow Rate: 180 ml/min (for non-intuabed patients)
CO2 (Carbon Dioxide):
Pulse Oximeter
SpO2 (Oxygen Saturation)
Pulse Rate:
Display:
Graphic Trend/Histogram:
Physical:
Electrical: